Aumeascon AM-D-pH Water RS485 Digital Online pH Sensor

 AM-D-pH  Digital Online pH Sensor

 


Aumeascon AM-D-pH digital pH electrode independently developed and produced by Aumeascon. The housing of pH sensor is SS316 for harsh application, easy to install, and it has high measurement accuracy, sensitive response, and can work stably for a long time. Built-in temperature probe, instant temperature compensation. Strong anti-interference ability, the longest output cable can reach 500 meters. It can be set and calibrated remotely, and the operation is simple. It can be widely used in monitoring the pH of thermal power, chemical fertilizer, metallurgy, environmental protection, pharmaceutical, biochemical, food, fermentation, brewing and tap water and other fields.

 

Technical Specification

Ø Standard RS485 Modbus.

Ø Maximum transmission distance 500 meters.

Ø Independent chip, anti-interference, strong stability.

Ø SS316 housing for harsh waste water.

 

Technical Parameters

Model

AM-D-pH

Measuring range

0~14pH  0~65

Accuracy

±0.1pH  ±0.5℃

Resolution

0.01pH  0.1℃

Power Supply

9~36V DC

Communication

RS485 Modbus RTU

Housing material

SS316 for fresh water ; Titanium for sea water

Connection

Upper G1

 

RS485 Modbus RTU Protocol

Address

Meaning

Range

Default value

Rate

R/W

Code

Indicate

0

Temp

0-1270

 

0.1℃

R

 

 

1

pH

0-1400

 

0.01pH

R

 

 

2

Manual Temp

0-1270

250

0.1℃

R

 

 

3

Measuring potential

-4141~4141

 

0.1mV

R

 

 

5

Temp state

0-2

 

 

R

 

=0: normal;

=1: too high or too low

=2: no sensor connected

8

Address

1-254

2

 

R/W

 

Change the slave address

9

Baud rate

4800-19200

9600

BPS

R/W

 

Only allow: 4800,9600,19200

10

Parameter Recovery

 

 

 

W

1996

revert to default

11

Reset the Instrument

1524

 

 

W

1524

Restart the instrument

13

Instrument Calibration

 

 

 

W

1

Zero potential calibration: The electrode is placed in a buffer of pH 7.00 or 6.86.

2

Slope calibration: electrode is placed in pH4.00 (10.00) or

in buffer at pH 4.00 (9.18). Buffer type changed by register 20

14

Zero point adjustment

-5916~5916

0

0.01mV

R/W

 

When register 13 is written with instruction 1, the value is automatically updated

15

Slope

0-100

100

1%

R

 

When register 13 is written with instruction 2, the value is automatically updated

20

Standard Solution

0,1

0

 

R/W

 

0:Standard Solution 4.00,7.00,10.00;

1:Standard Solution 4.00,6.86,9.18

 

Communication Format Details

Temperature data read command:

Address + function code + register start address + number of registers read + CRC check code (hexadecimal)

For example Tx:01 03 00 00 00 01 84 0A

Address

Code

Register start address

number of registers

CRC check code

01

03

0000

0001

840A

 

Temperature data return command:

Address + function code + data length + data + CRC check code (hexadecimal)

For example Rx:01 03 02 00 DF F9 DC

Address

Code

Data length

Temperature

CRC check code

01

03

02

00DF

F9DC

 

pH data return command:

Address + function code + data length + data + CRC check code (hexadecimal)

For example Rx: 01 03 02 02 AE 38 98

Address

Code

Register start address

number of registers

CRC check code

01

03

0001

0001

D5CA

 

pH data return command:

Address + function code + data length + data + CRC check code (hexadecimal)

For example Rx: 01 03 02 02 AE 38 98

Address

Code

Register start address

number of registers

CRC check code

01

03

02

02AE

3898

Convert the hexadecimal number 2AE to decimal by using the calculator (programmer mode) to get the value 686. The actual value of pH contains 2 decimal places, then the actual value is 686×0.01= 6.86

 

Wiring instructions

V+

V-

M-A

M-B

9~36V Positive

9~36V Negative

RS485-A

RS485-B

 

Contact info:

Shanghai Aumeascon Automation Technology Company 

Website:www.aumeascon.com 

Mob/WA:86-18721238342 

Email: Sales@aumeascon.com

Address:No. 95, Shihu New Road, Shihudang Town, Songjiang District, Shanghai.China , Zip code:201604 

 

 

 

评论

此博客中的热门博文

Aumeascon AM-MP Water Quality Multi-parameters Meter

Aumeascon AM-D-GEC Digital Graphite Conductivity Electrode

Aumeascon AM-D-NH3N Digital Ammonia Nitrogen Sensor